Investigating the robustness of the Anzai respiratory gating system

dc.contributor.authorAnderson, Maggie (Margaret)
dc.date.accessioned2014-01-13T20:42:49Z
dc.date.available2014-01-13T20:42:49Z
dc.date.issued2013en
dc.description.abstractThis research was undertaken in order to investigate the robustness of the Anzai respiratory gating system. Tests were performed to verify the transfer of image data, to identify the method of gating and the accuracy of phase identification. It was found to have sizeable limitations which could result in either incorrectly gated images or serious artefacts. For these reasons it is recommended it be used under the guidance of a suitably qualified physicist.en
dc.identifier.urihttp://hdl.handle.net/10092/8796
dc.identifier.urihttp://dx.doi.org/10.26021/8861
dc.language.isoen
dc.publisherUniversity of Canterbury. Physicsen
dc.relation.isreferencedbyNZCUen
dc.rightsCopyright Maggie (Margaret) Andersonen
dc.rights.urihttps://canterbury.libguides.com/rights/thesesen
dc.subjectRespiratory gatingen
dc.subjectCTen
dc.titleInvestigating the robustness of the Anzai respiratory gating systemen
dc.typeTheses / Dissertations
thesis.degree.disciplineMedical Physicsen
thesis.degree.grantorUniversity of Canterburyen
thesis.degree.levelMastersen
thesis.degree.nameMaster of Scienceen
uc.bibnumber1974705en
uc.collegeFaculty of Scienceen