University of Canterbury Home
    • Admin
    UC Research Repository
    UC Library
    JavaScript is disabled for your browser. Some features of this site may not work without it.
    View Item 
    1. UC Home
    2. Library
    3. UC Research Repository
    4. Faculty of Science | Te Kaupeka Pūtaiao
    5. Science: Research Data
    6. View Item
    1. UC Home
    2.  > 
    3. Library
    4.  > 
    5. UC Research Repository
    6.  > 
    7. Faculty of Science | Te Kaupeka Pūtaiao
    8.  > 
    9. Science: Research Data
    10.  > 
    11. View Item

    MARS spectral molecular imaging of lamb tissue: data collection and image analysis (2013-10-31)

    Thumbnail
    View/Open
    Data_for_Soft_tissue_paper.zip (4.608Gb)
    Type of Content
    Datasets
    UC Permalink
    http://hdl.handle.net/10092/8531
    
    Collections
    • Science: Research Data [4]
    Authors
    Aamir, R.
    Chernoglazov, A.
    Bateman, C. J.
    Butler, A.P.H.
    Butler, P.H.
    Anderson, N.G.
    Bell, S.T.
    Panta, R.
    Healy, J.L.
    Mohr, J.L.
    Rajendran, K.
    Walsh, M.F.
    de Ruiter, J.A.
    Giesig, S.P.
    Woodfield, T.
    Renaud, P.F.
    Brooke, L.V.
    Majid, S.A.
    Clyne, R.
    Glendinning, R.
    Bones, P.J.
    Billinghurst, Mark cc
    Bartneck, Christoph cc
    Mandalika, H.
    Grasset, R.
    Schleich, N.
    Scott, N.
    Nik, S. J.
    Opie, A.
    Janmale, T.
    Tang, D.N.
    Kim, D.
    Doesburg, R.M.
    Zainon, R.
    Ronaldson, J.P.
    Cook, N.J.
    Smithies, D.
    Hodge, K.
    show all
    Abstract

    In this experiment, a meat specimen was prepared from a fresh lamb chop, which included muscle (water-like), fat (lipid-like) and bone (calcium-like) regions, and scanned. We used a 2 mm thick CdTe sensor (128×128), bump bonded at 110 µm to Medipix3RX readout chip and installed in the MARS-CT4 scanner. In charge summing mode (CSM), 720 circular projections over 360° were acquired using a Source-Ray SB-80-1K x-ray tube (Source-Ray Inc, Ronkonkoma, NY) with a tungsten anode having 1.8-mm-Al equivalent intrinsic filtration. Several vertical positions of the CdTe Medipix3RX camera were used to create a virtual detector to cover the 23 mm field of view (FOV). The source to detector distance (SDD) was 131.8mm and object to detector distance (ODD) was 48mm. Camera readout was performed using the MARS readout system. Before the measurements, threshold equalization with respect to the noise edge, and energy calibration of the detector, were performed. Flat-field measurements (500 flat-fields per energy bin) were taken before specimen scanning to correct for variations in pixel response. Dark-field images (50 dark-fields per energy bin) were also acquired before and after the scan. The tube was operated at 50 kVp with a current of 120 μA and using four low energy thresholds (15, 20, 25, 30 keV). The exposure time of each acquisition was 40 ms. For data analysis and HU calibration, a phantom having CaCl2 (320 mg/ml) and lipid (vegetable oil), along with air and water, was also scanned with the same parameters mentioned earlier. The raw data in DICOM format from the scanner was processed using the MARS Image Processing Suite. More detail can be found from the relevant paper data paper.

    Citation
    Aamir, R., Chernoglazov, A., Bateman, C. J., Butler, A. P. H., Butler, P. H., Anderson, N. G., … Hodge, K. (2013). MARS spectral molecular imaging of lamb tissue: data collection and image analysis. arXiv:1311.4528 [physics]. Retrieved from http://arxiv.org/abs/1311.4528
    This citation is automatically generated and may be unreliable. Use as a guide only.

    Related items

    Showing items related by title, author, creator and subject.

    • Reducing beam hardening effects and metal artefacts using Medipix3RX: With applications from biomaterial science 

      Rajendran, K.; Walsh, M.F.; de Ruiter, N.J.A.; Chernoglazov, A.I.; Panta, R.K.; Butler, A.P.H.; Butler, P.H.; Bell, S.T.; Anderson, N.G.; Woodfield, T.B.F.; Tredinnick, S.J.; Healy, J.L.; Bateman, C.J.; Aamir, R.; Doesburg, R.M.N.; Renaud, P.F.; Gieseg, S.P.; Smithies, D.J.; Mohr, J.L.; Mandalika, V.B.H.; Opie, A.M.T.; Cook, N.J.; Ronaldson, J.P.; Nik, S.J.; Atharifard, A.; Clyne, M.; Bones, P.J.; Bartneck, Christoph; Grasset, R.; Schleich, N.; Billinghurst, Mark (University of Canterbury. Biological SciencesUniversity of Canterbury. Computer Science and Software EngineeringUniversity of Canterbury. Electrical and Computer EngineeringUniversity of Canterbury. Human Interface Technology LaboratoryUniversity of Canterbury. Mathematics and StatisticsUniversity of Canterbury. Mechanical EngineeringUniversity of Canterbury. Physics and AstronomyUniversity of Canterbury. Biomolecular Interaction Centre, 2014)
      This paper discusses methods for reducing beam hardening effects using spectral data for biomaterial applications. A small-animal spectral scanner operating in the diagnostic energy range was used. We investigate the use ...
    • Reducing beam hardening effects and metal artefacts in spectral CT using Medipix3RX 

      Rajendran, K; Walsh, M. F.; de Ruiter, N. J. A.; Chernoglazov, A. I.; Panta, R. K.; Butler, P. H.; Bell, S. T.; Woodfield, T. B. F.; Tredinnick, J.; Healy, J. L.; Bateman, C. J.; Aamir, R.; Doesburg, R. M. N.; Renaud, P. F.; Gieseg, S. P.; Smithies, D. J.; Mohr, J. L.; Mandalika, V. B. H.; Opie, A. M. T.; Cook, N. J.; Ronaldson, J. P.; Nik, S. J.; Atharifard, A.; Clyne, M.; Bones, P. J.; Bartneck, Christoph; Grasset, R.; Schleich, N.; Billinghurst, Mark; Butler, A. P. H.; Anderson, N. G. (2014-02-05)
      This paper discusses methods for reducing beam hardening effects and metal artefacts using spectral x-ray information in biomaterial samples. A small-animal spectral scanner was operated in the 15 to 80 keV x-ray energy ...
    • Reducing beam hardening effects and metal artefacts in spectral CT using Medipix3RX 

      Rajendran, K; Walsh, M. F.; de Ruiter, N. J. A.; Chernoglazov, A. I.; Panta, R. K.; Butler, P. H.; Bell, S. T.; Woodfield, T. B. F.; Tredinnick, J.; Healy, J. L.; Bateman, C. J.; Aamir, R.; Doesburg, R. M. N.; Renaud, P. F.; Gieseg, S. P.; Smithies, D. J.; Mohr, J. L.; Mandalika, V. B. H.; Opie, A. M. T.; Cook, N. J.; Ronaldson, J. P.; Nik, S. J.; Atharifard, A.; Clyne, M.; Bones, P. J.; Bartneck, Christoph; Grasset, R.; Schleich, N.; Billinghurst, Mark; Butler, A. P. H.; Anderson, N. G. (2013-11-21)
      Studies on beam hardening and metal artefact reduction using Medipix All Resolution System (MARS) spectral scanner were carried out. Four datasets are provided - titanium phantom, titanium scaffold, magnesium scaffold and ...
    Advanced Search

    Browse

    All of the RepositoryCommunities & CollectionsBy Issue DateAuthorsTitlesSubjectsThesis DisciplineThis CollectionBy Issue DateAuthorsTitlesSubjectsThesis Discipline

    Statistics

    View Usage Statistics
    • SUBMISSIONS
    • Research Outputs
    • UC Theses
    • CONTACTS
    • Send Feedback
    • +64 3 369 3853
    • ucresearchrepository@canterbury.ac.nz
    • ABOUT
    • UC Research Repository Guide
    • Copyright and Disclaimer
    • SUBMISSIONS
    • Research Outputs
    • UC Theses
    • CONTACTS
    • Send Feedback
    • +64 3 369 3853
    • ucresearchrepository@canterbury.ac.nz
    • ABOUT
    • UC Research Repository Guide
    • Copyright and Disclaimer