| Title: | Data entry errors and design for model-based tight glycemic control in critical care |
| Authors: | Ward, L. Steel, J. LeCompte, A.J. Evans, A. Tan, C.S. Penning, S. Shaw, G.M. Desaive, T. Chase, J.G. |
| Keywords: | critical care data entry error glycemic control human factors ICU intensive care intensive insulin therapy STAR TGC user interface |
| Issue Date: | 2012 |
| Citation: | Ward, L., Steel, J., LeCompte, A.J., Evans, A., Tan, C.S., Penning, S., Shaw, G.M., Desaive, T., Chase, J.G. (2012) Data entry errors and design for model-based tight glycemic control in critical care. Journal of Diabetes Science and Technology, 6(1), pp. 135-143. |
| Abstract: | Introduction: Tight glycemic control (TGC) has shown benefits but has been difficult to achieve consistently. Model-based methods and computerized protocols offer the opportunity to improve TGC quality but require human data entry, particularly of blood glucose (BG) values, which can be significantly prone to error. This study presents the design and optimization of data entry methods to minimize error for a computerized and model-based TGC method prior to pilot clinical trials.
Method: To minimize data entry error, two tests were carried out to optimize a method with errors less than the 5%-plus reported in other studies. Four initial methods were tested on 40 subjects in random order, and the best two were tested more rigorously on 34 subjects. The tests measured entry speed and accuracy. Errors were reported as corrected and uncorrected errors, with the sum comprising a total error rate. The first set of tests used randomly selected values, while the second set used the same values for all subjects to allow comparisons
across users and direct assessment of the magnitude of errors. These research tests were approved by the University of Canterbury Ethics Committee.
Results: The final data entry method tested reduced errors to less than 1–2%, a 60–80% reduction from reported values. The magnitude of errors was clinically significant and was typically by 10.0 mmol/liter or an order of magnitude but only for extreme values of BG < 2.0 mmol/liter or BG > 15.0–20.0 mmol/liter, both of which
could be easily corrected with automated checking of extreme values for safety.
Conclusions: The data entry method selected significantly reduced data entry errors in the limited design tests presented, and is in use on a clinical pilot TGC study. The overall approach and testing methods are easily performed and generalizable to other applications and protocols. |
| Publisher: | University of Canterbury. Mechanical Engineering |
| Research Fields: | Field of Research::11 - Medical and Health Sciences::1103 - Clinical Sciences::110310 - Intensive Care Field of Research::11 - Medical and Health Sciences::1102 - Cardiovascular Medicine and Haematology::110202 - Haematology Field of Research::11 - Medical and Health Sciences::1103 - Clinical Sciences::110306 - Endocrinology Field of Research::09 - Engineering::0903 - Biomedical Engineering |
| URI: | http://hdl.handle.net/10092/7239 |
| Rights URI: | http://library.canterbury.ac.nz/ir/rights.shtml |
| Appears in Collections: | Journal Articles
|
Items in UC Research Repository are protected by copyright, with all rights reserved, unless otherwise indicated.
|